Health product recall

Drill Sleeve Guide (2019-08-22)

Starting date:
August 22, 2019
Posting date:
October 11, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71255


Last updated:

Affected Products

Drill Sleeve Guide

Reason

On August 22, 2019, Ad-Tech Medical Instrument Corporation decided to voluntarily recall all drill sleeve guides which are intended to be used only with the 2.4mm diameter cranial drill bit. This recall has been initiated due to an investigation being performed that has identified potential issues with both of our raw material suppliers.

The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery. Although these devices do not make patient contact and are not expected to result in death, this issue could result in a delay in procedure.

Affected products

Drill Sleeve Guide

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • DSG-6.3-090-2.4N
  • DSG-90-2.3N
Companies
Manufacturer

AD-TECH MEDICAL INSTRUMENT CORPORATION

400 West Oakview Parkway

Oak Creek

53154

Wisconsin

UNITED STATES