Health product recall

Dragonfly Opstar Imaging Catheter

Brand(s)
Last updated

Summary

Product
Dragonfly Opstar Imaging Catheter
Issue
Medical devices - Quality
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected Products

Lot or Serial Number

Model or catalog number

Dragonfly Opstar Imaging Catheter

8257087, 8220675, 8257085, 8220673, 8274128

1014651

Issue

The proximal marker on devices from these lots may separate from the device. A dislodged marker may require additional intervention, including unplanned additional coronary intervention, or surgery. To date Abbott has received 5 complaints related to this issue, one which resulted in an embolism which required an additional intervention to resolve with no immediate adverse consequences. Potential risks include embolism, cardiovascular injury and myocardial ischemia.

Recall start date: Apr 11, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Abbott Medical

4 Robbins Road, Westford, Massachusetts

United States, 01886

Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type I
Identification number
RA-64111