Health product recall

Dräger Jaundice Meter JM-103 and JM-105 (2018-05-09)

Starting date:
May 9, 2018
Posting date:
May 18, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66804

Affected products

A. DRAGER JAUNDICE METER JM-103

B. DRAGER JAUNDICE METER JM-105

Reason

Draeger's continuous post market surveillance has shown that some users have concerns about the method used by JM-103 & JM-105 to indicate that a taken measurement is out of range (higher than the measuring range of the device). As described in the instructions for use and in the device's training materials, the JM-103 displays a blinking '- - -' and the JM-105 displays a blinking '- o -' when the measurement is out of range. The measuring range is defined to 340 mol/l / 20 mg/dl.

Affected products

A. DRAGER JAUNDICE METER JM-103

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

MU20606

Companies
Manufacturer

Draeger Medical Systems, Inc.

3135 Quarry Road

Telford

18969

UNITED STATES



B. DRAGER JAUNDICE METER JM-105

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

MU20105

Companies
Manufacturer

Draeger Medical Systems, Inc.

3135 Quarry Road

Telford

18969

UNITED STATES