Health product recall

DISPOSABLE BIPOLAR ABLATOR, ASPIRATING & NON-ASPIRATING (2021-02-01)



Starting date:
February 1, 2021
Posting date:
March 11, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75093


Last updated:

Affected Products 

A. DISPOSABLE BIPOLAR ABLATOR, ASPIRATING

B. DISPOSABLE BIPOLAR ABLATOR, NON-ASPIRATING

Reason

Package integrity failure leading to potential breach of the sterile barrier and potential contamination of the sterile product inside packaging. There can be a potential breach of the sterile packaging barrier due to the development of pinholes at the crease of the double-folded edge of the current package configuration. Issue was discovered after worst case transportation / handling / shipping/extreme temperature conditions showed package integrity failure in 10.5% of samples tested.

Affected products

A. DISPOSABLE BIPOLAR ABLATOR, ASPIRATING

Lot or serial number

SN# C3350A LOT# 0517D

SN# C3390A LOT# 0517D

Model or catalog number

C3350A

C3390A

Companies
Manufacturer

Bovie Medical Corporation

5115 Ulmerton Road

Clearwater

33760

Florida

UNITED STATES



B. DISPOSABLE BIPOLAR ABLATOR, NON-ASPIRATING

Lot or serial number

SN# C1860NA LOT# 0817H

SN# C2455NA LOT# 0817H

SN# C3350NA LOT# 0817H

SN# C3390NA LOT# 0817H

Model or catalog number

C1860NA

C2455NA

C3350NA

C3390NA

Companies
Manufacturer

Bovie Medical Corporation

5115 Ulmerton Road

Clearwater

33760

Florida