Discovery PET/CT 600 and 690 Systems - Main Units (2013-12-06)
- Starting date:
- December 6, 2013
- Posting date:
- January 13, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Hospitals, Healthcare Professionals
- Identification number:
- RA-37463
Recalled products
- Discovery PET/CT 600 System - Main Unit
- Discover Pet/Ct 690 System - Main Unit
Reason
A software issue exists when using list mode replay, such that subsequent reconstructions with that data could contribute to an image with scatter over-estimation and overcorrection. Reconstructed images with this issue will have many fewer counts than the acquired scan data and a high scatter fraction. This issue may affect studies that use list mode replay and are reconstructed after subsequent acquisitions.
Affected products
A. Discovery PET/CT 600 System - Main Unit
Lot or serial number
Not applicable
Model or catalog number
- S9116KC
Companies
- Manufacturer
-
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
B. Discover Pet/Ct 690 System - Main Unit
Lot or serial number
Not applicable
Model or catalog number
- S9164JC
Companies
- Manufacturer
-
GE Medical Systems LLC
3200 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES