Health product recall

Discovery PET/CT 600 and 690 Systems - Main Units (2013-12-06)

Starting date:
December 6, 2013
Posting date:
January 13, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-37463

Recalled products

  1. Discovery PET/CT 600 System - Main Unit
  2. Discover Pet/Ct 690 System - Main Unit

Reason

A software issue exists when using list mode replay, such that subsequent reconstructions with that data could contribute to an image with scatter over-estimation and overcorrection. Reconstructed images with this issue will have many fewer counts than the acquired scan data and a high scatter fraction. This issue may affect studies that use list mode replay and are reconstructed after subsequent acquisitions.

Affected products

A. Discovery PET/CT 600 System - Main Unit

Lot or serial number

Not applicable

Model or catalog number
  • S9116KC
Companies
Manufacturer

GE Medical Systems LLC

3200 North Grandview Blvd.

Waukesha

53188

Wisconsin

UNITED STATES



B. Discover Pet/Ct 690 System - Main Unit

Lot or serial number

Not applicable

Model or catalog number
  • S9164JC
Companies
Manufacturer

GE Medical Systems LLC

3200 North Grandview Blvd.

Waukesha

53188

Wisconsin

UNITED STATES