Health product recall

Dimension Xpand System - Xpand Plus Basic Analyzer (Class 2 and 3), Xpand Plus with HM Analyzer (Class 2 and 3) and Dimension EXL 200 Analyzer (2018-01-16)

Starting date:
January 16, 2018
Posting date:
February 1, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65876

Affected products

  1. DIMENSION XPAND SYSTEM - XPAND PLUS BASIC ANALYZER
  2. DIMENSION XPAND PLUS BASIC ANALYZER - class 2
  3. DIMENSION EXL 200 ANALYZER
  4. DIMENSION XPAND SYSTEM - XPAND PLUS WITH HM ANALYZER - class 2
  5. DIMENSION XPAND SYSTEM - XPAND PLUS WITH HM ANALYZER

Reason

Manufacturer identified an issue with Dimension Software version 10.2.1. The software version does not permit the user to enter or edit calibration coefficients for applicable assays without support from a Siemens customer service representative.

Affected products

A. DIMENSION XPAND SYSTEM - XPAND PLUS BASIC ANALYZER

Lot or serial number

2004081160

Model or catalog number

765000.921

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark

19714

Delaware

UNITED STATES



B. DIMENSION XPAND PLUS BASIC ANALYZER - class 2

Lot or serial number

2004081160

Model or catalog number

765000.921

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark

19714

Delaware

UNITED STATES



C. DIMENSION EXL 200 ANALYZER

Lot or serial number

DE270253

Model or catalog number

778041.911

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark

19714

Delaware

UNITED STATES



D. DIMENSION XPAND SYSTEM - XPAND PLUS with hm analyzer - class 2

Lot or serial number

2001081906

Model or catalog number

765000.931

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark

19714

Delaware

UNITED STATES



E. DIMENSION XPAND SYSTEM - XPAND PLUS WITH HM ANALYZER

Lot or serial number

2001081906

Model or catalog number

765000.931

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark

19714

Delaware

UNITED STATES