Dimension Vista Systems Aliquot Plates (2013-10-30)
- Starting date:
- October 30, 2013
- Posting date:
- January 6, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-37287
Recalled products
- Dimension Vista Systems Aliquot Plates
Reason
Manufacturer has confirmed intermittent Aliquot plate barcode failure to read errors causing the plate to be discarded. The error is generated when the Aliquot plate is being loaded onboard the instrument or recalled from the Aliquot elevator for an add-on test. If the Aliquot plate cannot be read, it is discarded into the waste. Not all plates within a lot exhibit the failure. Not all instruments experience failures with the same lot of Aliquot plates.
Affected products
A. Dimension Vista Systems Aliquot Plates
Lot or serial number
- N2809513
- N2810813
- N2811413
- N2817013
Model or catalog number
- 10445210
- KS840
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES