Dimension Vista System Prot 1 Con and Cal (2017-05-31)

Starting date:
May 31, 2017
Posting date:
July 6, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63802

Affected Products

A. Dimension Vista System Prot1 Con
B. Dimension Vista System Prot1 Cal

Reason

Siemens Healthineers has confirmed a negative bias of approximately -20% for results obtained with Dimension Vista CER Flex Reagent Cartridge when calibrating CER with affected PROT 1 CAL lots. PROT 1 controls for CER are similarly biased.

Affected products

A. Dimension Vista System Prot1 Con

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

KC715
KC716