Dimension Vista System ? High Sensitivity Troponin I (TNIH) (2020-07-27)
- Starting date:
- July 27, 2020
- Posting date:
- August 28, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73767
Last updated:
2020-08-28
Affected Products
Dimension Vista System ? High Sensitivity Troponin I (TNIH)
Reason
Siemens has observed a negative bias across the analytical measurement range of the TnIH assay. The average bias observed for patient samples using the dimension vista TnIH lots 20008bb, 20035BC, and 20135BB when compared with a control TnIH lot was -20% for lot 20008bb and -23% for lot 20135BB. A maximum bias of -26% in patient samples around the 99th percentile with these specific lots was observed. Quality control will not detect the negative patient bias. TnIH lot 20008bb and lot 20035BC contain the same reagents therefore, performance of the two lots is equivalent.
Affected products
Dimension Vista System ? High Sensitivity Troponin I (TNIH)
Lot or serial number
- 20008bb
- 20035BC
- 20135BB
Model or catalog number
K6427
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.