Health product recall

Dimension Vista System ? High Sensitivity Troponin I (TNIH) (2020-07-27)

Starting date:
July 27, 2020
Posting date:
August 28, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73767



Last updated:
2020-08-28

Affected Products

Dimension Vista System ? High Sensitivity Troponin I (TNIH)

Reason

Siemens has observed a negative bias across the analytical measurement range of the TnIH assay. The average bias observed for patient samples using the dimension vista TnIH lots 20008bb, 20035BC, and 20135BB when compared with a control TnIH lot was -20% for lot 20008bb and -23% for lot 20135BB. A maximum bias of -26% in patient samples around the 99th percentile with these specific lots was observed. Quality control will not detect the negative patient bias. TnIH lot 20008bb and lot 20035BC contain the same reagents therefore, performance of the two lots is equivalent.

Affected products

Dimension Vista System ? High Sensitivity Troponin I (TNIH)

Lot or serial number
  • 20008bb
  • 20035BC
  • 20135BB
Model or catalog number

K6427

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, P.O. Box 6101

Newark

19714

Delaware

UNITED STATES