Dimension Vista System - Hemoglobin A1c (HbA1c), Dimension EXL system - Hemoglobin A1c (Hb1c),Dimension Xpand system - Hemoglobin A1c (Hb1c) and Dimension Cinical Chemistry System - Hemoglobin a (2014-09-24)
- Starting date:
- September 24, 2014
- Posting date:
- October 10, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-41745
Recalled Products
A) Dimension Vista System - Hemoglobin A1c (HbA1c)
B) Dimension EXL system - Hemoglobin A1c (Hb1c)
C) Dimension Xpand system - Hemoglobin A1c (Hb1c)
D) Dimension Cinical Chemistry System - Hemoglobin a
Reason
Siemens Healthcare Diagnostics is conducting a field recall for Dimension Vista HbA1c and Dimension Hb1C flex reagent lots listed in section 64(a) of this report. Siemens has confirmed that these reagent cartridge lots exhibit positive bias averaging 0.4% [4.4 mmol/mol] Hemoglobin A1c units and occasionally up to 1.0% [11 mmol/mol] HbA1c units for patient samples when compared to the national glycol-hemoglobin standardization program (NGSP). QC samples may exhibit a similar bias. As the customer base may differ based on the platform being utilized, Siemens is issuing two Urgent Field Safety Notices (UFSN): one UFSN is being issued for the affected lots of Dimension Vista Hemoglobin A1c (HbA1c). The second UFSN is being issued for the affected lots of Dimension Hemoglobin A1c (HbA1c). Customers will receive the letters pertinent to them. Both letters contain the same content, differing only in the specified product code and lot numbers.
Affected products
A) Dimension Vista System - Hemoglobin A1c (HbA1c)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
10470481
K3105A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
B) Dimension EXL system - Hemoglobin A1c (Hb1c)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
10483822
DF105A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
C) Dimension Xpand system - Hemoglobin A1c (Hb1c)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
10483822
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
D) Dimension Cinical Chemistry System - Hemoglobin a
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
10483822
DF105A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES