Health product recall

DIMENSION VISTA SYSTEM - CHEM CALIBRATORS (2016-03-01)

Starting date:
March 1, 2016
Posting date:
March 11, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57486

Affected Products

DIMENSION VISTA SYSTEM - CHEM CALIBRATORS

Reason

Siemens Healthcare Diagnostics has received complaints of "calibrator insert missing" alerts on Dimension Vista Chemistry 3 Calibrator lot 5md004. Internal investigation has confirmed that the vial barcode label reads KC130 instead of the correct catalog number KC130A. When this mismatch occurs between the vial barcode label and the Instructions For Use (IFU) label, the "calibrator insert missing" error is triggered and displayed on the Dimension Vista instrument screen. There is no risk of erroneous results when this issue occurs as no results are generated.

Affected products

DIMENSION VISTA SYSTEM - CHEM CALIBRATORS

Lot or serial number
  • 5MD004
Model or catalog number
  • KC130A
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark

19714-6101

Alabama

UNITED STATES