Dimension vista System-Chem calibrators

Starting date:
September 20, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60454

Affected Products

Reason

Siemens Healthcare Diagnostics has confirmed that DIMENSION VISTA CHEM 1 CAL Lots 5GM081 and 5GM082 may produce depressed results with current lots of calcium reagent. Although the affected CHEM 1 CAL lots are expired, the calcium calibration interval is 90 days, therefore it is possible that calibrations using the calibrator lots are active.

Affected products

Dimension vista System-Chem calibrators

Lot or serial number
  • 5GM081
  • 5GM082
Model or catalog number
  • K110B
  • KC110
Companies
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES