This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Dimension vista System-Chem calibrators

Starting date:
September 20, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60454

Affected Products

Reason

Siemens Healthcare Diagnostics has confirmed that DIMENSION VISTA CHEM 1 CAL Lots 5GM081 and 5GM082 may produce depressed results with current lots of calcium reagent. Although the affected CHEM 1 CAL lots are expired, the calcium calibration interval is 90 days, therefore it is possible that calibrations using the calibrator lots are active.

Affected products

Dimension vista System-Chem calibrators

Lot or serial number
  • 5GM081
  • 5GM082
Model or catalog number
  • K110B
  • KC110
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark

19714-6101

Delaware

UNITED STATES