Health product recall

Dimension Vista Lipase (LIPL), Dimension Clinical Chemistry, EXL and Xpand System - Lipase Flex (LIPL) Assay (2018-09-13)

Starting date:
September 13, 2018
Posting date:
October 5, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67882

Affected products

A. Dimension Vista Lipase (LIPL)

B. Dimension Clinical Chemistry System -Lipase Flex (LIPL) Assay

C. Dimension Xpand System - Lipase Flex (LIPL) Assay

D. Dimension EXL System - Lipase (LIPL) Assay

Reason

Siemens Healthcare Diagnostics has confirmed lot-to-lot variability for QC and patient results when using the Dimension LIPL lots GD9002, GA9003, GB9032, GA9038 and Dimension Vista LIPL lots 18128BA, 18172BA, 18186BB. Not all laboratories will observe this issue.

The lot-to-lot bias observed for patient samples using the affected Dimension Lots when compared with a control lot could be as high as 29%. The lot-to-lot bias observed for patient samples using the affected Dimension Vista Lots when compared with a control lot ranged from 9% to 34%.

Affected products

A. Dimension Vista Lipase (LIPL)

Lot or serial number

18128BA

18186BB

Model or catalog number

K3056

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark

19714

Delaware

UNITED STATES


B. Dimension Clinical Chemistry System -Lipase Flex (LIPL) Assay

Lot or serial number

GA9003

GA9038

GB9032

GD9002

Model or catalog number

DC56

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark

19714

Delaware

UNITED STATES


C. Dimension Xpand System - Lipase Flex (LIPL) Assay

Lot or serial number

GA9003

GA9038

GB9032

GD9002

Model or catalog number

DC56

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark

19714

Delaware

UNITED STATES


D. Dimension EXL System - Lipase (LIPL) Assay

Lot or serial number

GA9003

GA9038

GB9032

GD9002

Model or catalog number

DC56

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark

19714

Delaware

UNITED STATES