Health product recall

DIMENSION VISTA AND EMIT II OPIATE ASSAYS

Starting date:
August 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60076

Affected Products

  1. DIMENSION VISTA SYSTEM - URINE OPIATES SCREEN
  2. DIMENSION CLINICAL CHEMISTRY SYSTEM - OPIATES (UOPI)
  3. EMIT II PLUS OPIATES ASSAY

Reason

Siemens investigation confirmed that the introduction of a new SYVA polyclonal antibody lot (G1) for use in the dimension urine opiates screen has resulted in an alteration of specificity for Levallorphan, Levorphanol, Nalorphine and Oxycodone at the 300 and 2000 NG/ml cutoff concentration. As a result in the change in sensitivity to the cross reactant, the claims for these compounds are no longer accurate for either the 300 or 2000 NG/ml cutoff as stated within the instructions for use (IFU). Further investigation confirmed that SYVA, Beckman Coulter and Advia Chemistry Opiate reagents are impacted as well due to a common raw material used in the assays.

Affected products

A. DIMENSION VISTA SYSTEM - URINE OPIATES SCREEN

Lot or serial number
  • 15078BB
  • 15168BA
  • 15252AF
  • 15308BA
  • 16019BA
  • 16090BB
Model or catalog number
  • K5093
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark, Delaware

19714-6101

UNITED STATES



B. DIMENSION CLINICAL CHEMISTRY SYSTEM - OPIATES (UOPI)

Lot or serial number
  • EA6091
  • FA6258
  • FA7069
  • GA6324
  • GB6275
  • GB7029
  • GC6160
Model or catalog number
  • DF93A
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark, Delaware

19714-6101

UNITED STATES



C. EMIT II PLUS OPIATES ASSAY

Lot or serial number
  • H2
  • H3
  • H4
  • H5
  • H6
  • J1
  • J2
Model or catalog number
  • 9B309UL
  • 9B329UL
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101

Newark, Delaware

19714-6101

UNITED STATES