Dimension Systems-Ammonia Assays (2017-03-07)
- Starting date:
- March 7, 2017
- Posting date:
- March 24, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-62788
Affected Products
A. Dimension Exl System-Ammonia (Amm) Assay
B. Dimension Xpand System-Ammonia (Amm) Assay
C. Dimension Exl System-Ammonia (Amm) Assay
Reason
In September 2014, Siemens Healthcare Diagnostics issued customer notification 14-62 that addressed a high rate of "abnormal assay" test report messages for its Dimension Ammonia (AMM) Assay. With this letter, Siemens is reducing the AMM open well stability from 3 days to 1 day (24 hours) and providing instructions for managing this change. Dimension AMM instructions for use (IFU) revision E will reflect this change.
The new version of dimension software, 10.2, will do the following:
- automatically move to a new well every day. This decrease in the open well maximum duration reduces the amount of time the AMM reagent is exposed to any potential contaminates, reduces the need for customers to manually move to a new reagent well, and aids in the prevention of erroneous flagging.
- contain the below mean factor limit of 0.90.
Siemens is working to make Dimension Software10.2 available to all customers as quickly as possible.
Affected products
A. Dimension Exl System-Ammonia (Amm) Assay
Lot or serial number
All lots
Model or catalog number
10711991
DF119
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
B. Dimension Xpand System-Ammonia (Amm) Assay
Lot or serial number
All lots
Model or catalog number
10711991
DF119
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
C. Dimension Exl System-Ammonia (Amm) Assay
Lot or serial number
All lots
Model or catalog number
10711991
DF119
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES