Dimension Reaction Vessels (2013-12-19)
- Starting date:
- December 19, 2013
- Posting date:
- January 24, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-37647
Recalled products
- Dimension Reaction Vessels
Reason
Siemens Healthcare Diagnostic has observed an increase in the frequency of complaints for vessel feeder errors. The errors observed include "670: vessel feeder track sensor stuck on full", and "749: vessel track empty". Complaints have been confirmed with several Dimension reaction vessel lots. A low percentage of reaction vessels from those lots have been observed to have flashing on the flange of the vessel that may jam in the vessel track and generate an error.
Affected products
A. Dimension Reaction Vessels
Lot or serial number
All lots
Model or catalog number
- RXV1A
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES