This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Dimension Products – Total Iron Binding Capacity (IBCT)

Starting date:
December 21, 2012
Posting date:
January 7, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-16671

Recalled Products

A. Dimension Clinical Chemistry – IBCT
B. Dimension Xpand System – IBCT
C. Dimension EXL System – IBCT

Reason

Through monitoring of method Performance Siemens has determined that the analytical sensitivity claim of 6 mcg/dl (1.1 mcmol/l) stated in the instructions for use is no longer valid for Dimension Total Iron Binding Capacity (ibct), and is modifying that claim to 36 mcg/dl (6.44 mcmol/l).

Affected products

A. Dimension Clinical Chemistry – IBCT

Lot or serial number

More than 10 Numbers, Contact manufacturer

Model or catalog number

DF84

Companies
Manufacturer
Siemens Healthcare Diagnostics Inc.

B. Dimension Xpand System - IBCT

Lot or serial number

More than 10 Numbers, Contact manufacturer

Model or catalog number

DF84

Companies
Manufacturer
Siemens Healthcare Diagnostics Inc.

C. Dimension EXL System – IBCT

Lot or serial number

More than 10 Numbers, Contact manufacturer

Model or catalog number

DF84

Companies
Manufacturer
Siemens Healthcare Diagnostics Inc.