Health product recall

Dimension EXL System-(Troponin) TNI Assay (2013-09-12)

Starting date:
September 12, 2013
Posting date:
October 28, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36443

Recalled Products

Dimension EXL System-(Troponin) TNI Assay 

Reason

Some vials of Dimension Loci Cardiac Troponin I calibrator under-recover Troponin I, resulting in over-recovery of QC and patient samples following calibration. Testing on patient samples demonstrated an average upward shift of 24% (Range 1%-33%) when compared to an unaffected lot.

Affected products

Dimension EXL System-(Troponin) TNI Assay 

Lot or serial number
  • 3DD007
Model or catalog number
  • 10464336
  • RC621
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, PO Box 6101

Newark

19714-6101

Delaware

UNITED STATES