Dimension Exl System - Cardiac Troponin-I (CTnI) Assay
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Dimension Exl System - Cardiac Troponin-I (CTnI) Assay |
GA2323 |
RF421C |
Issue
Siemens Healthcare Diagnostics Inc. has confirmed that Dimension Cardiac Troponin I (lTni and CTnI) lots (ea2314 and ga2323) shows a positive bias with lithium heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/ml. Troponin values as high as 0.20 ng/ml have been observed with lithium heparin plasma that are expected to be within the 99th percentile. Serum samples collected from normal, healthy individuals whose results fall within the reference interval or below the 99th percentile of 0.07 ng/ml were not impacted by this issue.
Recall start date: May 24, 2022
Additional information
Details
Siemens Healthcare Diagnostics Inc.
500 Gbc Drive, Newark, Delaware, United States, 19714
Get notified
Receive notifications for new and updated recalls and alerts by category.