Health product recall

DIMENSION CLINICAL CHEMISTRY, DIMENSION XPAND, DIMENSION EXL SYSTEMS (2019-07-19)

Starting date:
July 19, 2019
Posting date:
August 2, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70597


Last updated:

Affected Products

A. DIMENSION CLINICAL CHEMISTRY SYS. - FERRITIN (FERR) FLEX REAGENT CARTRIDGE

B. DIMENSION XPAND SYSTEM - FERRITIN (FERR) ASSAY

C. DIMENSION EXL SYSTEM - FERRITIN (FERR) ASSAY

Reason

Siemens Healthcare Diagnostics has confirmed that the Dimension® Ferritin (FERR) Flex® reagent lot listed in Table 1 may exhibit a negative bias during the open well stability time period. Siemens has determined that FERR values can potentially decrease by -50% at a concentration of 22 ng/mL [μg/L] and -48% at a concentration of 285 ng/mL [μg/L] after a 10 day open well period. Siemens has also determined that FERR values can potentially decrease by approximately -15% after a 3 day open well period. Quality control and patient samples are affected similarly.

Affected products

A. DIMENSION CLINICAL CHEMISTRY SYS. - FERRITIN (FERR) FLEX REAGENT CARTRIDGE

Lot or serial number

EA9227

Model or catalog number

RF440

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, P.O. Box 6101

Newark

19714

Delaware

UNITED STATES



B. DIMENSION XPAND SYSTEM - FERRITIN (FERR) ASSAY

Lot or serial number

EA9227

Model or catalog number

10444946

RF440

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, P.O. Box 6101

Newark

19714

Delaware

UNITED STATES



C. DIMENSION EXL SYSTEM - FERRITIN (FERR) ASSAY

Lot or serial number

EA9227

Model or catalog number

10444946

RF440

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

500 GBC Drive, Mailstop 514, P.O. Box 6101

Newark

19714

Delaware

UNITED STATES