Dimension Ammonia Assays (2016-12-23)
- Starting date:
- December 23, 2016
- Posting date:
- January 18, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-61832
Affected Products
- Dimension Vista System - Ammonia Assay
- Dimension EXL System - Ammonia Assay
- Dimension Xpand System - Ammonia Assay
- Dimension Clinical Chemistry - AMM Assay
Reason
Siemens has determined that five (5) lots of Dimension Ammonia Assay (DF119) and three (3) lots of Dimension Vista Ammonia Assay (K3119) Flex Reagent Cartridges do not meet the 60-day calibration interval claim due to reagent instability and may show an abnormal assay message. These lots may exhibit accuracy shifts for patient and/or quality control results. This may cause laboratories to recalibrate more frequently than the 60-day claim in the IFU.
Affected products
A. Dimension Vista System - Ammonia Assay
Lot or serial number
- 16187BE
- 16225BB
- 16265AB
Model or catalog number
- 10711992
- K3119
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
B. Dimension EXL System - Ammonia Assay
Lot or serial number
- BA7194
- BA7250
- EA7223
- EB7180
- FB7152
Model or catalog number
- 10711991
- DF119
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
C. Dimension Xpand System - Ammonia Assay
Lot or serial number
- BA7194
- BA7250
- EA7223
- EB7180
- FB7152
Model or catalog number
- 10711991
- DF119
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES
D. Dimension Clinical Chemistry - AMM Assay
Lot or serial number
- BA7194
- BA7250
- EA7223
- EB7180
- FB7152
Model or catalog number
- 10711991
- DF119
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714-6101
Delaware
UNITED STATES