Digital Diagnost C90
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Digital Diagnost C90 |
Not applicable. |
9890 010 83905 |
Issue
Operators may see an incorrect orientation of images during the first exam upon system start. due to an issue in the firmware of the wallstand vs2 board, the system will rotate the amplimat field selection by 90 degrees. The anatomic position markers will become malpositioned and could potentially be associated with the opposite side of the anatomy. To date, Philips has not received reports of patient, user, or bystander harm in relation to these potential issues. affected units are not required to be removed from service and are safe for continued use consistent with the device instructions for use and this medical device recall letter.
Recall start date: June 2, 2022
Additional information
Details
Philips Medical Systems DMC GmbH
Roentgenstrasse 24, Hamburg, Germany, 22335