This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Dialog+ SW 9.XX (2015-11-18)

Starting date:
November 18, 2015
Posting date:
January 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56560

Affected Products

A. Dialog+ SW 9.XX

Reason

Potential problem if servicing device and interchanging tubing between valve block level regulation and diaphragm pump if the functionality of the level regulation system is not tested as requested in the service manual. An interchanged tubing results in an inverse function of the manual level regulation system.

Affected products

A. Dialog+ SW 9.XX

Lot or serial number

All lots.

Model or catalog number

710200CV

Companies
Manufacturer

B. Braun Avitum Ag

Schwarzenberger Weg 73-79

Melsungen

34212

GERMANY