This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

DGDE- DIRECTGUIDE DRILL EXTENDER (2015-10-23)

Starting date:
October 23, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56292

Affected Products

DGDE- DIRECTGUIDE DRILL EXTENDER

Reason

Implant Direct Sybron Manufacturing LLC is recalling 3 lots of the DGDE Direct Guide Drill Extender (DGDE) and 22 lots of associated products that contain DGDE as an accessory.  Through internal Regulatory Affairs evaluation (HHE) process, the company found a possible safety issue with the retention of the drill when the DGDE is in use. This discrepancy may lead to the possibility of the DGDE dropping the drill.  The company's current data indicates the possibility of this occurring is unlikely (0.69%).

Affected products

DGDE- DIRECTGUIDE DRILL EXTENDER (2015-10-23)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • BST
  • CST
  • DGDE
  • IBST
  • ICST
  • RBST
  • RCST
  • SPCST
Companies
Manufacturer

Implant Direct Sybron Manufacturing LLC

3050 East Hillcrest Drive

Thousand Oaks

91362

California

UNITED STATES