Devon Light Glove Kit (2016-12-23)
- Starting date:
- December 23, 2016
- Posting date:
- February 14, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Hospitals, Healthcare Professionals
- Identification number:
- RA-62172
Affected Products
- Devon Birthday Kit C-Section Basic
- Devon Light Glove
- Devon OR Mini Kit
Reason
Customers have reported that, on rare occasion, the Devon Light Glove may split upon application to the Devon Light Handle Adapter. Some of the reported splits resulted from difficult application of the Light Glove to the Handle Adapter. More recently, clinicians have reported finding splits in the Light Glove following surgery completion, where no difficulty in application of the light glove was encountered or finding splits directly out of the package. A split in the Light Glove causes a breach in the sterile field and can increase the potential for infection. Medtronic has received notice of two patient adverse events (infection) in which the Light Glove splits were found at the conclusion of surgery.
Affected products
A. Devon Birthday Kit C-Section Basic
Lot or serial number
All lots with 630xxx and lower
Model or catalog number
31404851
Companies
- Manufacturer
-
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
B. Devon Light Glove
Lot or serial number
All lots 630xxx and below
Model or catalog number
- 31140208
- 31140216
- 31140257
Companies
- Manufacturer
-
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
C. Devon OR Mini Kit
Lot or serial number
All lots 630xxx and below
Model or catalog number
- 31141552
- 50000148
Companies
- Manufacturer
-
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES