DELFIA / AutoDELFIA PIGF 1-2-3
Brand(s)
Summary
Contact the importer REVVITY HEALTH SCIENCES CANADA, INC. at TURKUQMRESPONSE@REVVITY.COM.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
DELFIA / AutoDELFIA PIGF 1-2-3 |
More than 10 numbers, contact manufacturer. |
B055-301 |
Issue
We have become aware through customer complaints that the measured PlGF analyte concentrations in part of the patient samples (approximately 15%) may be decreased ≥ 50%. The decreased PlGF concentration results may cause raise in the amount of high risk/false positive pre-eclampsia and Down's syndrome screening results.
Recall Start Date: May 29, 2024
Additional information
Details
Wallac Oy
Mustionkatu 6, Turku, Finland, 20750
Get notified
Receive emails about new and updated recall and safety alerts.