This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Dash and Tram 451 Patient Monitors and Patient Data Module (2013-02-26)

Starting date:
February 26, 2013
Posting date:
October 22, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55470

Affected Products

  • A. Dash 3000 Patient Monitor
  • B. Dash 5000 Patient Monitor
  • C. Patient Data Module
  • D. Patient Data Module
  • E. Dash 4000 Patient Monitor

Reason

ECG and respiration signal noise caused by Dash, PDM and/or Tram ECG cable connector movement may result in reduced ECG, respiration and arrhythmia detection performance.

Affected products

A. Dash 3000 Patient Monitor

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

DASH3-XXXX-XXXX-XXXX

Companies
Manufacturer

GE Medical Systems Information Technologies, Inc.

8200 West Tower Avenue

Milwaukee

Wisconsin

UNITED STATES



B. Dash 5000 Patient Monitor

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

DASH5-XXXX-XXXX-XXXX

Companies
Manufacturer

GE Medical Systems Information Technologies, Inc.

8200 West Tower Avenue

Milwaukee

Wisconsin

UNITED STATES



C. Tram 451 Patient Monitor

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

T451=A

Companies
Manufacturer

GE Medical Systems Information Technologies, Inc.

8200 West Tower Avenue

Milwaukee

Wisconsin

UNITED STATES



D. Patient Data Module

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

PDM-XXX-XXX-XXXX

Companies
Manufacturer

GE Medical Systems Information Technologies, Inc.

8200 West Tower Avenue

Milwaukee

Wisconsin

UNITED STATES



E. Moniteur patient Dash 4000

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

DASH4-XXXX-XXXX-XXXX

Companies
Manufacturer

GE Medical Systems Information Technologies, Inc.

8200 West Tower Avenue

Milwaukee

Wisconsin

UNITED STATES