Health product recall

Darco System 7.0 Headless Screw (Sterile)

Last updated

Summary

Product
Darco System 7.0 Headless Screw (Sterile)
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalogue number

Darco System 7.0 Headless Screw (Sterile)

1733433

DCS01660

Issue

Subsequent to a product complaint (April 8 2024), Stryker confirmed that one specific lot of Darco™ 7.0mm Headless Screws was manufactured without the hex feature required for mating with the associated screwdriver.
Recall start date: September 3, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies
Wright Medical Technology Inc.
1023 Cherry Road, Memphis, Tennessee, United States, 38117
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-76033

Get notified

Receive emails about new and updated recall and safety alerts.