Da Vinci Surgical Systems
Brand(s)
Summary
CAPA 552131 started to address issue. Customers provided with further instructions in response.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Da Vinci X Surgical System Is4200 |
Contact manufacturer. |
PS4200 |
Da Vinci Xi Surgical System Is4000 |
Contact manufacturer |
IS4000 |
Issue
An increase in reports in which a loose instrument carriage is observed on the universal side manipulator (usm) or the instrument arm of the Da Vinci X and Xi systems. The loose instrument carriage is an outcome of a linear rail component being out of specification. The loose instrument carriage makes it possible for the carriage to wobble along the linear rail in response to force applied to the carriage or to the shaft of an installed instrument.
Recall start date: February 23, 2023
Additional information
Details
Intuitive Surgical Inc.
1266 Kifer Road, Sunnyvale, California, United States, 94086
Get notified
Receive emails about new and updated recall and safety alerts.