CVM Calibration Verification Material (2019-05-16)
- Starting date:
- May 16, 2019
- Posting date:
- June 7, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70123
Last updated: 2019-06-07
Affected Products
CVM Calibration Verification Material
Reason
Siemens Healthcare Diagnostics has confirmed an error in the CVM material performance variability range pertaining to the level 1 pCO2 range for the siemens rapidlab 1260/1265 systems, where the correct CVM material performance variability range for pCO2 is 121.3-150.5 mmHg, instead of the published range of 135.1-164.3 mmHg. This is a value assignment issue for performance variability range not a product issue. Only customers with RAPIDLab 1260/1265 systems are affected.
Affected products
CVM Calibration Verification Material
Lot or serial number
350802
Model or catalog number
10316535
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.