Health product recall

Custom Tubing Packs (2021-03-19)

Starting date:
March 19, 2021
Posting date:
April 1, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-75283



Last updated:
2021-04-01

Affected Products 

  1. HEART LUNG PERFUSION PACK WITH SOFTLINE
  2. CUSTOM PACK WITH SOFTLINE
  3. PERFUSION SET WITH SOFTLINE COATING
  4. PERFUSION TUBING PACK WITH BIOLINE COATING
  5. PERFUSION SET WITH BIOLINE COATING

Reason

The packs listed above contain component 700000172 which is a bacterial filter. DEKRA Certification GmbH (MCP's notified body) issued the decision letter regarding the design dossier review of Customized Tubing Sets (CTPs) under MDD recertification. Due to the insufficient evidence regarding the shelf life of this component, the overall shelflife of the finished tubing packs containing this component must be reduced by one year.

Affected products

A. HEART LUNG PERFUSION PACK WITH SOFTLINE

Lot or serial number
  • 701062277
Model or catalog number
  • BO-HQV 48707
Companies
Manufacturer

MAQUET CARDIOPULMONARY GMBH

KEHLER STR. 31, RASTATT, GERMANY, 76437

RASTATT

76437

GERMANY


B. CUSTOM PACK WITH SOFTLINE

Lot or serial number
  • 701063761
Model or catalog number
  • BO-HQV 55004
Companies
Manufacturer

MAQUET CARDIOPULMONARY GMBH

KEHLER STR. 31, RASTATT, GERMANY, 76437

RASTATT

76437

GERMANY


C. PERFUSION SET WITH SOFTLINE COATING

Lot or serial number
  • 701070954
  • 701070955
Model or catalog number
  • BO-H 55815
  • BO-H 55816
Companies
Manufacturer

MAQUET CARDIOPULMONARY GMBH

KEHLER STR. 31, RASTATT, GERMANY, 76437

RASTATT

76437

GERMANY


D. PERFUSION TUBING PACK WITH BIOLINE COATING

Lot or serial number
  • 701065229
  • 701065246
  • 701066180
Model or catalog number
  • BEQ-H 48715
  • BEQ-H 48730
  • BEQ-H 84702
Companies
Manufacturer

MAQUET CARDIOPULMONARY GMBH

KEHLER STR. 31, RASTATT, GERMANY, 76437

RASTATT

76437

GERMANY


E. PERFUSION SET WITH BIOLINE COATING

Lot or serial number
  • 701062275
Model or catalog number
  • BEQ-HQV 48707
Companies
Manufacturer

MAQUET CARDIOPULMONARY GMBH

KEHLER STR. 31, RASTATT, GERMANY, 76437

RASTATT

76437

GERMANY