CUSTOM PACK WITH SOFTLINE (2021-03-23)
- Starting date:
- March 23, 2021
- Posting date:
- April 1, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75297
Last updated:
2021-04-01
Affected Products
CUSTOM PACK WITH SOFTLINE
Reason
The sterile pouch (component #700007026) is used to pack the accessories that are placed inside the primary sterile barrier (blister) and intended to organize the articles inside the primary packaging. The pouch #700007026 is used for packing the centrifugal pump ""rotaflow"" and the manifold line,
which are not pre-connected to the tubing set.
The ability of the sterile pouch to maintain its integrity as a sterile barrier during storage and transport is insufficiently demonstrated. The sterility of the device itself is however not affected as the integrity of the primary sterile barrier (blister) is not compromised by storage and transport conditions.
Affected products
CUSTOM PACK WITH SOFTLINE
Lot or serial number
- 3000146594
- 3000147939
- 92299428
Model or catalog number
- BO-HQV 48412
Companies
- Manufacturer
-
MAQUET CARDIOPULMONARY GMBH
KEHLER STR. 31, RASTATT, GERMANY, 76437
RASTATT
76437
GERMANY