Health product recall

CUSTOM PACK WITH SOFTLINE (2021-03-23)



Starting date:
March 23, 2021
Posting date:
April 1, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75297


Last updated:

Affected Products

CUSTOM PACK WITH SOFTLINE

Reason

The sterile pouch (component #700007026) is used to pack the accessories that are placed inside the primary sterile barrier (blister) and intended to organize the articles inside the primary packaging. The pouch #700007026 is used for packing the centrifugal pump ""rotaflow"" and the manifold line,

which are not pre-connected to the tubing set.

The ability of the sterile pouch to maintain its integrity as a sterile barrier during storage and transport is insufficiently demonstrated. The sterility of the device itself is however not affected as the integrity of the primary sterile barrier (blister) is not compromised by storage and transport conditions.

Affected products

CUSTOM PACK WITH SOFTLINE

Lot or serial number
  • 3000146594
  • 3000147939
  • 92299428
Model or catalog number
  • BO-HQV 48412
Companies
Manufacturer

MAQUET CARDIOPULMONARY GMBH

KEHLER STR. 31, RASTATT, GERMANY, 76437

RASTATT

76437

GERMANY