This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

CS 8100 and CS 8100 3D (2015-01-27)

Starting date:
January 27, 2015
Posting date:
February 6, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43633

Recalled products

  1. CS 8100
  2. CS 8100 3D

Reason

Carestream has identified an issue related to the manufacturing process of CS 8100 and CS 8100 3D tomographic digital panoramic x-ray systems, manufactured between June 2012 and September 2014, which could result in the head unit descending unexpectedly and contacting the patient. The issue was found at an installation and nobody was injured. It has been determined that the root cause was a circlip (retaining clip) on a pin used to mate the lower and upper portions of the stand not being fitted at the time of manufacture. This allowed the locating pin to move out of position within the location hole during transportation. There is no hazard during transport and the issue would now be found at any new installation. Testing has shown that even without both circlips fitted, the pin does not fall out during use, but could move, therefore to ensure all installed devices are not affected and will perform as designed through their expected life, a field inspection recall is being initiated.

Affected products

A. CS 8100

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 530 3045
  • 531 1329
Companies
Manufacturer
Carestream Health Inc.
150 Verona Street
Rochester
14608
New York
UNITED STATES

B. CS 8100 3D

Lot or serial number
  • CIID071
Model or catalog number
  • 5311162
Companies
Manufacturer
Carestream Health Inc.
150 Verona Street
Rochester
14608
New York
UNITED STATES