Creatine kinase-MB (CK-MB) (2018-08-05)
- Starting date:
- August 5, 2018
- Posting date:
- January 11, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68732
Affected products
Creatine kinase-MB (CK-MB)
Reason
Following internal investigation of customer complaints, the manufacturer confirmed that an increased absolute absorbance for both calibration standards and reduced control recoveries could occur for the product Creatine Kinase-MB (#05168562190) lot # 31195601 used on cobas c 701/702 module.
Affected products
Creatine kinase-MB (CK-MB)
Lot or serial number
31195601
Model or catalog number
05168562190
Companies
- Manufacturer
-
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY