Health product recall

Creatine kinase-MB (CK-MB) (2018-08-05)

Starting date:
August 5, 2018
Posting date:
January 11, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68732

Affected products

Creatine kinase-MB (CK-MB)

Reason

Following internal investigation of customer complaints, the manufacturer confirmed that an increased absolute absorbance for both calibration standards and reduced control recoveries could occur for the product Creatine Kinase-MB (#05168562190) lot # 31195601 used on cobas c 701/702 module.

Affected products

Creatine kinase-MB (CK-MB)

Lot or serial number

31195601

Model or catalog number

05168562190

Companies
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY