Creatine Kinase (CK)
Brand(s)
Summary
Contact the ROCHE CARE CENTER at 1-877-273-3433 or (450) 686-7111 and by e-mail at LAVAL.TECHINFO@ROCHE.COM.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Creatine Kinase (CK) |
755471 |
07190794190 |
Issue
Roche received complaints from US customers for Creatine Kinase (CK) lot 755471 (expired June-2024) on cobas c 311/501/502 and INTEGRA 400 plus about abnormal calibration signals, flagged calibrations (Sens.E) and control recovery issues. Subsequent investigation identified a positive bias at the lower measuring range and a negative bias at the higher measuring range. The issue affects a small number of cassettes of the mentioned lot.
Recall Start Date: August 19, 2024