This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Corometrics 170 Series Monitors (2015-05-04)

Starting date:
May 4, 2015
Posting date:
June 1, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53557

Recalled Products

  1. Corometrics 170 Series Monitors

Reason

Some of the Corometrics 171, 172, 173 and 174 series monitors were delivered with the FHR lower alarm limit configured incorrectly at 50 BPM (proper factory default setting is 120 BPM). A lower than expected FHR, if left undetected, could result in fetal bradycardia and oxygen deprivation to the fetus. There have been no injuries reported as a result of this issue.

Affected products

A. Corometrics 170 Series Monitors

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

Not applicable

Companies
Manufacturer

Wipro GE Healthcare Private Ltd.

No 4

Kadugodi Industrial Area

560067

INDIA