Health product recall

Corflo-Max PEG Kit Pull (2020-03-19)



Starting date:
March 19, 2020
Posting date:
April 17, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72803


Last updated:

Affected Products

Corflo-Max PEG Kit Pull

Reason

Avanos Medical (formerly Halyard Health) has received 3 complaint reports stating that the Corflo* PEG tube may be blocked at the center of the adapter (occluded). Avanos reviewed its inventory and demonstrated the defect rates to be 1.4% for full occlusion and 2.2% for partial occlusion, based on lots produced after the implementation of the suspected root cause. A PEG tube that is occluded may not be functional and the PEG procedure may be delayed or aborted. A new PEG tube placement may be required to continue with the procedure. There is no need for concern in patients who have had PEGs placed that function normally, which are de facto unoccluded.

Affected products

Corflo-Max PEG Kit Pull

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 30-6012
  • 30-6016
  • 30-6020
  • 50-6012
  • 50-6020
  • 50-6316
  • 50-6520
Companies
Manufacturer

Avanos Medical Inc.

5405 Windward Parkway

Alpharetta

30004-3894

Georgia

UNITED STATES