This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Coonrad/Morrey Elbow Replacement Set

Starting date:
November 19, 2012
Posting date:
December 10, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-16311

Recalled Products

A. Coonrad/Morrey Elbow Replacement Set

Reason

Zimmer is initiating a Lot specific recall of the Coonrad/Morrey Total Elbow Replacement Set bushings and pins for use with size regular elbow, due to the potential that the affected lot could include an extra small inner pin instead of the required regular inner pin.

An extra small inner pin is not long enough to fully engage a regular outer pin, therefore the click described in the surgical technique would not be heard or felt during intra-operative assembly.

Affected products

A. Coonrad/Morrey Elbow Replacement Set

Lot or serial number

61644802

Model or catalog number

32-8106-000-13

Companies
Manufacturer

Zimmer Inc.