Health product recall

COOK Vacuum Pump and Aspiration Unit (2018-03-01)

Starting date:
March 1, 2018
Posting date:
March 23, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66278

Affected products

A. VACUUM PUMP

B. COOK ASPIRATION UNIT

Reason

The internal mains wiring does not fully comply with the requirements of the medical electrical equipment standard IEC60601-1 Edition 3.0. IEC60601-1 states that conductors and connectors of medical electrical equipment shall be secured or insulated so that accidental detachment shall not result in a hazardous situation. The K-MAR-5200 mains wiring is secured, but the mechanism is not considered suitable under the current version of IEC60601-1.

Affected products

A. VACUUM PUMP

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

K-MAR-5200

Companies
Manufacturer

William A. Cook Australia, Pty. Ltd.

95 Brandl Street

Eight Mile Plains, Queensland

4113

AUSTRALIA



B. COOK ASPIRATION UNIT

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

K-MAR-5200-US

Companies
Manufacturer

William A. Cook Australia, Pty. Ltd.

95 Brandl Street

Eight Mile Plains, Queensland

4113

AUSTRALIA