Health product recall

CONTOUR NEXT CONTROL SOLUTION (2015-06-01)

Starting date:
June 1, 2015
Posting date:
July 7, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53982

Recalled Products

  1. CONTOUR NEXT CONTROL SOLUTION (NORMAL)
  2. CONTOUR NEXT CONTROL SOLUTION (HIGH)
  3. CONTOUR NEXT CONTROL SOLUTION (LOW)

Reason

  1. Control result not marked as a control test: Bayer Healthcare has received user reports on CONTOUR NEXT control test results that are higher than the control range published on the packaging of contour next test strips and not recognized as control tests by the meter.  All lots of CONTOUR NEXT control solution (levels low, normal and high) are affected.
  2. Control result outside of control range:  Another issue may affect any control solution lots when used with of CONTOUR NEXT test strip lots with expiration dates through November 2016.  Due to unforeseen material variation, lots of contour next control solution may give control test results that are outside of the assigned control range as published on the packaging of CONTOUR NEXT test strips.  These control results are properly marked as control tests by the meter therefore this issue does not impact insulin pumps.

Affected products

A. CONTOUR NEXT CONTROL SOLUTION (NORMAL)

Lot or serial number

All lots.

Model or catalog number
  • 7325
    Companies
    Manufacturer

    Bayer Healthcare LLC Subsidiary of Bayer Corporation

    P.O. Box 2004, 430 S. Bieger Street

    Mishawaka

    Indiana

    UNITED STATES



    B. CONTOUR NEXT CONTROL SOLUTION (HIGH)

    Lot or serial number

    All lots.

    Model or catalog number
    • 7324
    Companies
    Manufacturer

    Bayer Healthcare LLC Subsidiary of Bayer Corporation

    P.O. Box 2004, 430 S. Bieger Street

    Mishawaka

    Indiana

    UNITED STATES



    C. CONTOUR NEXT CONTROL SOLUTION (LOW)

    Lot or serial number

    All lots.

    Model or catalog number
    • 7323
    Companies
    Manufacturer

    Bayer Healthcare LLC Subsidiary of Bayer Corporation

    P.O. Box 2004, 430 S. Bieger Street

    Mishawaka

    Indiana

    UNITED STATES