Health product recall

Contender Post-Op Knee Brace (2015-12-07)

Starting date:
December 7, 2015
Posting date:
January 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56770

Affected Products

  1. Contender Post-Op Knee Brace

Reason

Corflex received notification from a material supplier that a component used to manufacture the brace was out of specification. The hardness of the drop lock teeth could be insufficient, causing this feature to fail under stress. In circumstances where a sudden and/or higher than normal force is exerted against an engaged drop lock, the brace could fail to maintain the locked position and a loss of balance or undesired flexion/extension may occur at the affected knee joint, posing a safety risk to the patient.

Affected products

A. Contender Post-Op Knee Brace

Lot or serial number

all lots

Model or catalog number
  • 75-7500-000
  • 75-7500-011
  • 75-7500-038
  • 75-7500-075
  • 75-7500-087
  • 75-7500-231
  • 75-7550-000
Companies
Manufacturer

Corflex

669 E. INDUSTRIAL PARK DRIVE

MANCHESTER

03109

New Hampshire

UNITED STATES