Health product recall

ComboLab System - Main Unit, CardioLab System - Main Unit, Mac-Lab System - Main Unit, and Mac-Lab Systems - Integrated Electronics Box (2018-06-14)

Starting date:
June 14, 2018
Posting date:
July 27, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67298

Affected products

A. ComboLab System  - Main Unit

B. CardioLab System - Main Unit

C. Mac-Lab System - Main Unit

D. Mac-Lab Systems - Integrated Electronics Box

Reason

The following issue is being addressed by GE healthcare file no. FMI 12269: GE Healthcare has recently become aware that the system side of the power cord may break and expose the electrical conductors leading to the risk of shock. This can lead to injuries up to and including cardiac arrhythmia or cardiac arrest.

Affected products

A. ComboLab System  - Main Unit

Lot or serial number

SKP16397103TA

Model or catalog number

2092616-004

Companies
Manufacturer

GE Medical Systems Information Technologies, Inc.

8200 W Tower Ave

Milwaukee

53223

Wisconsin

UNITED STATES



B. CardioLab System - Main Unit

Lot or serial number

SKP17248451TA

SKP17258483TA

SKP17258484TA

Model or catalog number

2092616-002

Companies
Manufacturer

GE Medical Systems Information Technologies, Inc.

8200 W Tower Ave

Milwaukee

53223

Wisconsin

UNITED STATES



C. Mac-Lab System - Main Unit

Lot or serial number

SKP17218256TA

SKP17258477TA

SKP17258481TA

SKP17258485TA

SKP17258486TA

Model or catalog number

2092616-001

Companies
Manufacturer

GE Medical Systems Information Technologies, Inc.

8200 W Tower Ave

Milwaukee

53223

Wisconsin

UNITED STATES



D. Mac-Lab Systems - Integrated Electronics Box

Lot or serial number

SG411061097GA

SPH17358709TA

SPH18049347TA

Model or catalog number

2098841-001 (2098804-001)

P1008CV (2033852-011)

Companies
Manufacturer

GE Medical Systems Information Technologies, Inc.

8200 W Tower Ave

Milwaukee

53223

Wisconsin

UNITED STATES