CombiDiagnost R90
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
CombiDiagnost R90 |
All lots. |
709030 |
Issue
If the CombiDiagnost R90 System has a damaged wire in the System Power Distribution Unit (SPDU) within the m-cabinet and is connected to the ground, or the system experiences a power surge, there is the potential for a short circuit of the Electromagnetic Interference (EMI) filter or UA/UB board. If this issue occurs, the user may experience one or more of the following:
- electrical fire within the main-cabinet (m-cabinet)
- see or smell smoke
- hear a popping sound coming from the m-cabinet
Recall start date: October 28, 2021
Additional information
Details
Philips Medical Systems DMC GmbH
Roentgenstrasse 24, Hamburg, Germany, 22335