Health product recall

Cocaine Metabolite Micro-Plate EIA (2018-01-16)

Starting date:
January 16, 2018
Posting date:
February 9, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65924

Affected products

Cocaine Metabolite Micro-Plate EIA

Reason

Cocaine microplate kit failed the six month stability time point at initial testing, as well as the double operator repeat testing. A non-conformance report was initiated. The product may produce false positive patient results. The product provides only a preliminary analytical test result. A more specific alternative method should be used in order to obtain a confirmed analytical result.

Affected products

Cocaine Metabolite Micro-Plate EIA

Lot or serial number

6661251

Model or catalog number

1122S(B,C)

Companies
Manufacturer
Orasure Technologies, Inc.
220 East First Street
Bethlehem
18015-1360
Minnesota
UNITED STATES