This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

COBAS S201 System Pooling and Data Management (PDM) V3.0.35 (2015-04-20)

Starting date:
April 20, 2015
Posting date:
May 21, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53427

Recalled Products

A. COBAS S201 System Pooling and Data Management (PDM)

Reason

The pooling software can create pools containing more or less donors than originally intended during simultaneous pooling runs for both the Cobas TaqScreen MPX and Cobas TaqScreen WNV tests, when the donor tube carrier on the Hamilton STAR/STARlet pipettor is missing a donor tube in any position within the donor tube carrier.

Affected products

A. COBAS S201 System Pooling and Data Management (PDM)

Lot or serial number

All lots.

Model or catalog number

6660975190

Companies
Manufacturer

Roche Molecular Systems

Inc

1080 US Highway 202

New Jersey

UNITED STATES