This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

cobas IT middleware (2014-11-25)

Starting date:
November 25, 2014
Posting date:
December 5, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42765

Recalled Products

cobas IT middleware

Reason

There is an issue on cobas IT middleware where the masking of tests function (block release) is not working for modules if the master instrument is selected, therefore a remote probability that a delay or incorrect initial diagnosis may occur.

Affected products

cobas IT middleware

Lot or serial number

version 1.02.00 and above

Model or catalog number

06498027001
06498035001

Companies
Manufacturer
Roche Diagnostics GmbH
Sandhofer Strasse 116
Mannheim
68305
GERMANY