Health product recall

COBAS INTEGRA TINA-QUANT SOLUBLE TRANSFERRIN RECE (2015-07-03)

Starting date:
July 3, 2015
Posting date:
August 10, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54532

Affected Products

A. COBAS INTEGRA TINA-QUANT SOLUBLE TRANSFERRIN RECE

Reason

Roche received a customer complaint regarding elevated imprecision and too high recovery of STFR control set level 1 and DUP alarms during calibration with the reagent TQ Soluble Transferrin Receptor (STFR). The issue of elevated imprecision was confirmed by internal investigations on lot #606449 on Cobas C501 showed elevated imprecision of up to 8.9%. The other current reagent lots are not affected by this issue

Affected products

A. COBAS INTEGRA TINA-QUANT SOLUBLE TRANSFERRIN RECE

Lot or serial number

60644901

Model or catalog number

20763454122

Companies
Manufacturer

Roche Diagnostics Gmbh

Sandhoferstrasse 116

Mannheim

68305

GERMANY