This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Cobas E 411 Immunoassay Analyzer

Starting date:
July 7, 2017
Posting date:
August 14, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals
Identification number:
RA-64118

Reason

A potential software limitation in the Cobas E411 Immunoassay Analyzer could lead to a potential data mismatch under specific conditions.

Affected products

Cobas E 411 Immunoassay Analyzer

Lot or serial number

all software versions

Model or catalog number

  • 04775201001
  • 04775279001

Companies

Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY