This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

COBAS C 111 and COBAS C 111 WITH ISE STAND ALONE CONFIGURATION (2016-10-05)

Starting date:
October 5, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Hospitals, General Public, Healthcare Professionals
Identification number:
RA-60684

Affected Products

A. COBAS C 111 STAND ALONE CONFIGURATION

B. COBAS C 111 WITH ISE STAND ALONE CONFIGURATION

Reason

This recall is related to a situation that due to a software limitation, a measurement timing error could occur under specific conditions while using the COBAS C111 (CAT #04528778001) OR THE COBAS C111 with ISE (ion-selective electrode) (CAT # 04777433001) Analyzers.

Affected products

A. COBAS C 111 STAND ALONE CONFIGURATION

Lot or serial number

Serial Number: 4234

Model or catalog number

04528778001

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



B. COBAS C 111 WITH ISE STAND ALONE CONFIGURATION

Lot or serial number

Serial numbers:

S/N:50484

S/N:51459

S/N:51572

S/N:60142

S/N:60192

S/N:60289

S/N:60493

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY