Cobas B221 - Reference Electrode (2014-01-31)

Starting date:
January 31, 2014
Posting date:
March 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38337

Recalled Products

Cobas B221 - Reference Electrode

Reason

The Cobas B221 reference electrode has a lifetime of 52 weeks but is not documented in the analyzer manual. Customers had not been sufficiently informed regarding this limitation. They may find themselves using aged electrodes leading to leakage.  Ultimately, this may cause a pH and a NA+ offset which may result by incorrect diagnosis and treament.

Affected products

Cobas B221 - Reference Electrode

Lot or serial number

All lots

Model or catalog number

3111873180